beautypg.com

Wholebodyangio, Wholebodyangio 414, Vascular – Siemens EMOTION 16-Jun User Manual

Page 414

background image

414

Vascular

WholeBodyAngio

Indications:

For CT Angio spiral studies of the whole body.

For SOMATOM Emotion 16-slice configuration:

A range of 80 cm will be done in 17.87 sec.

Emotion 16

WholeBody

2

nd

reconstr.

kV

110

Effective mAs/

Quality ref. mAs

90

Rotation Time

0.6 sec.

Acquisition

16 x 1.2 mm

Slice collimation

1.2 mm

Slice width

6.0 mm

1.5 mm

Feed/Rotation

28.8 mm

Pitch Factor

1.50

Increment

6.0 mm

1.0 mm

Kernel

B31s

B31s

CTDI

Vol

6.48 mGy

Effective dose

Male: 9.33 mSv
Female: 10.11 mSv