Children – Siemens EMOTION 16-Jun User Manual
Page 588
588
Children
* The conversion factor for a 8-week-old child, and a
scan range of 150 mm was used.
Emotion 16
NeonateBody
kV
80
Effective mAs/
Quality ref. mAs
35
Rotation time
0.6 sec.
Acquisition
16 x 1.2 mm
Slice collimation
1.2 mm
Slice width
5.0 mm
Feed/Rotation
28.8 mm
Pitch Factor
1.50
Increment
5.0 mm
Kernel
B41s
CTDI
Vol
1.02 mGy
Effective dose
Male: 0.96 mSv*
Female: 1.07 mSv*
Emotion 6
NeonateBody
kV
80
Effective mAs/
Quality ref. mAs
35
Rotation time
0.8/0.6 sec.
Acquisition
6 x 2.0 mm
Slice collimation
2.0 mm
Slice width
5.0 mm
Feed/Rotation
18.0 mm
Pitch Factor
1.50
Increment
5.0 mm
Kernel
B41s
CTDI
Vol
1.02 mGy
Effective dose
Male: 3.23 mSv*
Female: 4.12 mSv*