beautypg.com

C-spinevol, C-spinevol 300, Spine – Siemens EMOTION 16-Jun User Manual

Page 300

background image

300

Spine

C-SpineVol

Indications:

Spiral mode for cervical spine studies, e.g. prolapse,
degenerative changes, trauma, tumors etc.

Three recon jobs are predefined for reconstruction: the
first for soft tissue axial, the second for soft tissue sag-
ittal and third for sagittal bone studies in 3D images
display view.

For SOMATOM Emotion 16-slice configuration:

A range of 15 cm will be covered in 21.53 sec.

For SOMATOM Emotion 6-slice configuration:

A range of 15 cm will be covered in 31.41 sec.

Emotion 16

C-Spine

2

nd

recon.

3

rd

recon.

kV

130

Effective mAs/
Quality ref. mAs

150

Rotation Time

1.0 sec.

Acquisition

16 x 0.6 mm

Slice collimation

0.6 mm

Slice width

2.0 mm

2.0 mm

2.0 mm

Feed/Rotation

7.7 mm

Pitch Factor

0.80

Increment

2.0 mm

2.0 mm

2.0 mm

Kernel

B31s

B20s

B60s

CTDI

Vol

18.75 mGy

Effective dose

Male: 3.49 mSv

Female: 3.71 mSv