C-spinevol, C-spinevol 300, Spine – Siemens EMOTION 16-Jun User Manual
Page 300
300
Spine
C-SpineVol
Indications:
Spiral mode for cervical spine studies, e.g. prolapse,
degenerative changes, trauma, tumors etc.
Three recon jobs are predefined for reconstruction: the
first for soft tissue axial, the second for soft tissue sag-
ittal and third for sagittal bone studies in 3D images
display view.
For SOMATOM Emotion 16-slice configuration:
A range of 15 cm will be covered in 21.53 sec.
For SOMATOM Emotion 6-slice configuration:
A range of 15 cm will be covered in 31.41 sec.
Emotion 16
C-Spine
2
nd
recon.
3
rd
recon.
kV
130
Effective mAs/
Quality ref. mAs
150
Rotation Time
1.0 sec.
Acquisition
16 x 0.6 mm
Slice collimation
0.6 mm
Slice width
2.0 mm
2.0 mm
2.0 mm
Feed/Rotation
7.7 mm
Pitch Factor
0.80
Increment
2.0 mm
2.0 mm
2.0 mm
Kernel
B31s
B20s
B60s
CTDI
Vol
18.75 mGy
Effective dose
Male: 3.49 mSv
Female: 3.71 mSv