Resp, Resp 502, Respiratory gating – Siemens EMOTION 16-Jun User Manual
Page 502
502
Respiratory Gating
Resp
Indications:
A spiral mode for patients with a respiration rate per
minute greater than 10 using respiration gating.
For SOMATOM Emotion 16-slice configuration:
A scan range of 30 cm will be covered in 94.95 sec.
For SOMATOM Emotion 6-slice configuration:
A scan range of 30 cm will be covered in 86.04 sec.
Emotion 16
Resp
2
nd
reconstr.
kV
130
Effective mAs/
Quality ref. mAs
320
Rotation time
0.6 sec.
Acquisition
16 x 1.2 mm
Slice collimation
1.2 mm
Slice width
3.0 mm
3.0 mm
Feed/Rotation
1.9
Pitch Factor
0.1
Increment
2.1 mm
2.1 mm
Kernel
B31s
B31s
CTDI
vol
35.84 mGy