Specials – Siemens EMOTION 16-Jun User Manual
Page 445
Specials
445
Application Procedure
1.Load and scan a spiral protocol of the interested
body region.
2.Scroll through the images to define a target slice.
3.Click on Same TP under Table position, in the routine
card. Move the table to the desired table position.
4.Turn on the light marker on the Gantry to localize the
entry point, and then start patient preparation.
5.Open the Patient Model Dialog. Check the check-
box Append. Select one of the predefined CAREVi-
sion scan protocols under Specials protocols and
then click OK.
6.Select the HandCARE position on the scan card.
7.Click Load and the Cancel/Move to scan on the pre-
selected table position.
8.Press the footswitch to start the Fluoroscopy.