Thorangioecgseq, Thorangioecgseq 392, Vascular – Siemens EMOTION 16-Jun User Manual
Page 392
392
Vascular
ThorAngioECGSeq
Indications:
Sequential, ECG triggered study for location and extent
of aneurysms, dissection, and ruptures of the thoracic
aorta with typical scan range of 29.9 cm.
Hints
• CARE Bolus may be used to optimize the bolus tim-
ing.
• Set the ROI for monitoring scan in the aortic arch
with triggering threshold of 120 HU, or use manual
triggering.
Emotion 6
ThorAngioECG
kV
130
Effective mAs/
Quality ref. mAs
80
Rotation Time
0.8 sec.
Acquisition
6 x 1.0 mm
Slice collimation
1.0 mm
Slice width
1.0 mm
Feed/Scan
6.0 mm
Kernel
B31s
Temp. resolution
*
*. depends on heart rate
up to 400 ms
CTDI
Vol
10.16 mGy
Effective dose
Male: 6.92 mSv
Female: 9.00 mSv
Contrast medium IV injection
Start delay
10 – 30 sec.
Flow rate
2.5 ml/sec.
Total amount
80 ml