Vascular – Siemens EMOTION 16-Jun User Manual
Page 373
Vascular
373
Emotion 16
CarotidAngio
2
nd
reconstr.
kV
110
Effective mAs/
Quality ref. mAs
70
Rotation time
0.6 sec.
Acquisition
16 x 0.6 mm
Slice collimation
0.6 mm
Slice width
5.0 mm
1.0 mm
Feed/Rotation
14.4 mm
Pitch Factor
1.50
Increment
5.0 mm
0.7 mm
Kernel
B30s
B20s
CTDI
Vol
5.60 mGy
Effective dose
Male: 1.28 mSv
Female: 1.41 mSv
Emotion 6
CarotidAngio
2
nd
reconstr.
kV
110
Effective mAs/
Quality ref. mAs
70
Rotation time
0.8/0.6 sec.
Acquisition
6 x 1.0 mm
Slice collimation
1.0 mm
Slice width
5.0 mm
1.25 mm
Feed/Rotation
9.0 mm
Pitch Factor
1.50
Increment
5.0 mm
0.8 mm
Kernel
B31s
B20s
CTDI
Vol
5.81 mGy
Effective dose
Male: 0.57 mSv
Female: 0.62 mSv