Children – Siemens EMOTION 16-Jun User Manual
Page 531
Children
531
* The conversion factor for a 7-year-old child, and a
scan range of 45 mm was used.
Emotion 16
InnerEar
kV
110
Effective mAs/
Quality ref. mAs
50
Rotation time
1.0 sec.
Acquisition
16 x 0.6 mm
Slice collimation
0.6 mm
Slice width
0.75 mm
Feed/Rotation
7.7 mm
Pitch Factor
0.80
Increment
0.5 mm
Kernel
H90s
CTDI
Vol
8.90 mGy
Effective dose
Male: 0.27 mSv*
Female: 0.28 mSv*
Emotion 6
InnerEar
kV
110
Effective mAs/
Quality ref. mAs
90
Rotation time
1.0 sec.
Acquisition
6 x 1.0 mm
Slice collimation
1.0 mm
Slice width
1.25 mm
Feed/Rotation
5.1 mm
Pitch Factor
0.85
Increment
1.0 mm
Kernel
H90s
CTDI
Vol
16.56 mGy
Effective dose
Male: 0.13 mSv*
Female: 0.12 mSv*