Orbit, Orbit 198, Head – Siemens EMOTION 16-Jun User Manual
Page 198
198
Head
Orbit
Indications:
Spiral mode for orbital studies, e.g. fracture.
For SOMATOM Emotion 16-slice configuration:
A range of 50 mm will be covered in 8.51 sec.
For SOMATOM Emotion 6-slice configuration:
A range of 50 mm will be covered in 11.80 sec.
Emotion 16
Orbit
2
nd
reconstr.
kV
130
Effective mAs/
Quality ref. mAs
70
Rotation Time
1.0 sec.
Acquisition
16 x 0.6 mm
Slice collimation
0.6 mm
Slice width
5.0 mm
1.0 mm
Feed/Rotation
7.7 mm
Pitch Factor
0.80
Increment
5.0 mm
0.7 mm
Kernel
H70s
H70s
CTDI
Vol
18.62 mGy
Effective dose
Male: 0.50 mSv
Female: 0.54 mSv