Vascular

415

Hints

• CARE Bolus may be used to optimize the bolus tim-

ing.

• Set the ROI for monitoring scan with triggering

threshold of 120 HU, or use manual triggering.

• If Topo length 1024 mm is not long enough, you can

also choose the 1540 mm long Topogram.

• Position the patient as feet first. Bend the feet

together if necessary.

Contrast medium IV injection
Start delay

10 – 20 sec.

Flow rate

3.0 – 3.5 ml/sec.

Total amount

120 – 150 ml