Children – Siemens EMOTION 16-Jun User Manual
Page 575
Children
575
* The conversion factor for a 7-year-old child, and a
scan range of 60 mm was used.
Emotion 6
HeadAngio
2
nd
reconstr.
kV
80
Effective mAs/
Quality ref. mAs
80
Rotation time
0.8/0.6 sec.
Acquisition
6 x 1.0 mm
Slice collimation
1.0 mm
Slice width
2.0 mm
1.25 mm
Feed/Rotation
9.0 mm
Pitch Factor
1.50
Increment
2.0 mm
0.8 mm
Kernel
C30s
C30s
CTDI
Vol
6.40 mGy
Effective dose
Male: 0.16 mSv*
Female: 0.16 mSv*