Table 4-5 – AMETEK ASD Series User Manual
Page 73

Sorensen ASD Series
Programming: Digital Interface Control
M551177-01 Rev A
4-15
Table 4-5. Command Register Bits
Bit
Field
Value
Description
1
ON
0
1
Turns off the unit.
Turns on the unit.
2
RESET_FAULT
0
1
No action.
Resets any faults.
3
REMOTE_SNS_DISABLE
0
1
Enables remote voltage sensing.
Disables remote voltage sensing.
4
ANALOG_CURRENT
0
1
Analog voltage programming mode.
Analog current programming mode.
5
IMPEDANCE_MONITOR
0
1
Disables output impedance monitoring.
Enables output impedance monitoring.
6
MODBUS_TIMEOUT
0
1
Disables monitoring of Modbus activity.
Enables monitoring of Modbus activity.
7
FOATING_POINT_ENABLED
0
1
Fractional register values use 32-bit IQ fractional encoding.
Fractional register values use 32-bit floating point encoding.
8
FDR_ENABLED
0
1
Flight Data Recorder periodic update disabled.
Flight Data Recorder periodic update enabled.
9
RESET_ENERGY_METER
0
1
No action.
Resets the energy meter to zero.
10
ANALOG_VOLTAGE_OVERRIDE
0
1
Analog Interface voltage set via external current or voltage signal.
Analog Interface voltage input disabled.
11
ANALOG_CURRENT_OVERRIDE
0
1
Analog Interface current set via external current or voltage signal.
Analog interface voltage input disabled.
12
ANALOG_POWER_OVERRIDE
0
1
Analog Interface power set via external current or voltage signal.
Analog Interface power input disabled.
13
DIGITAL PROGRAMMING MODE
0
1
Setpoints are taken from analog programming inputs.
Setpoints are taken from digital interface.
14
PROCESS CONTROL MODE
0
1
Standard modes: voltage/current/power.
Process control is active, modes: voltage/current/process.
15
DISABLE ANALOG SATURATION
0
1
Disabled analog inputs (bits 10, 11, 12) saturate to max setpoint.
Disabled analog inputs take setpoint from digital programming.
16
Not used