beautypg.com

Coolant connections – AMETEK ASD Series User Manual

Page 25

background image

Sorensen ASD Series

Installation

M551177-01 Rev A

2-27

2.3.2 CHASSIS REMOVAL FROM RACK

The slides have a Front Disconnect Feature and lock at full extension.

To return the chassis back into the rack from full extension, depress the
flat steel spring inward (located on the slides) and push the chassis back.

To disconnect and remove the chassis from the rack, depress the flat steel
spring inward (located on the slides) and pull the chassis forward.

When the chassis is at full extension, the flat springs are located
approximately one (1) inch behind the front EIA RETMA rails. Access the
springs with a flat blade screwdriver or similar device to release from lock-
out or to remove the chassis from the rack.

The slides can be mounted to the chassis with this spring oriented on the
top or the bottom of the slide.

2.4

COOLANT CONNECTIONS

WARNING!
Avoid risk of electrical shock. Ensure no electrical connections are made
until connections to the fluid fittings are secure and dry. Ensure that no
moisture comes into contact with electrical components or connections.

Do not remove cover. Refer to qualified service personnel.

Figure 2-1. Rear Panel Coolant Connectors

2.4.1 WATER QUALITY

DO NOT use de-ionized water or distilled water.

The quality of the supplied facility cooling water must have the following
specifications:

Particulate: Supply filtered to 10 microns.

Hardness: 60 ppm maximum (calcium & magnesium carbonates)