CB (E6) 09/2013

Page 99/144

On unit rear:
Exterior Lemo socket (8) on the

rear of the unit

Interior:
LEMO socket (B) located at the

upper right corner of the rear

wall of the inner chamber

Supplied LEMO connector:
on external and internal side

Figure 23: Pin allocation (front view) of the Lemo sockets and plug

When introducing a heat load into the inner chamber by inserting electrical devices like shakers or rollers,

the ambient temperature of the incubator must not exceed 25 °C / 77 °F. Otherwise, temperature control

to 37 °C / 98.6°F cannot be assured. When operating electrical devices inside the incubator, always

check the desired incubation temperature

before introducing in any cells.

CAUTION

Excessive inner temperature > 37 °C / 98.6°F when operating electrical devices

inside the incubator.
Damage to the cell cultures.

When operating electrical devices inside the incubator: Ambient temperature ≤ 25 °C /

77 °F.

Before introducing any cells, check the correct ambient temperature.


15.7 BINDER Gas Supply Service – External bottle changer for CO

2

, N

2

or O

2

(option)

The external bottle changer permits automatic switching to a second gas cylinder as soon as the first

cylinder is empty. It can be used for a maximum of two CB units.

Instructions 7001-0196 delivered with the external bottle changer (Art. no. 8012-0408)

describe its installation and operation.

2

1

7

6

8

3

4

5

3

4

6

7

5

2

1

8

3

4

6

7

5

2

1

8