Caution, Or o, Option) – BINDER CB 60 User Manual
Page 99
CB (E6) 09/2013
Page 99/144
On unit rear:
Exterior Lemo socket (8) on the
rear of the unit
Interior:
LEMO socket (B) located at the
upper right corner of the rear
wall of the inner chamber
Supplied LEMO connector:
on external and internal side
Figure 23: Pin allocation (front view) of the Lemo sockets and plug
When introducing a heat load into the inner chamber by inserting electrical devices like shakers or rollers,
the ambient temperature of the incubator must not exceed 25 °C / 77 °F. Otherwise, temperature control
to 37 °C / 98.6°F cannot be assured. When operating electrical devices inside the incubator, always
check the desired incubation temperature
before introducing in any cells.
CAUTION
Excessive inner temperature > 37 °C / 98.6°F when operating electrical devices
inside the incubator.
Damage to the cell cultures.
When operating electrical devices inside the incubator: Ambient temperature ≤ 25 °C /
77 °F.
Before introducing any cells, check the correct ambient temperature.
15.7 BINDER Gas Supply Service – External bottle changer for CO
2
, N
2
or O
2
(option)
The external bottle changer permits automatic switching to a second gas cylinder as soon as the first
cylinder is empty. It can be used for a maximum of two CB units.
Instructions 7001-0196 delivered with the external bottle changer (Art. no. 8012-0408)
describe its installation and operation.