This symbol on the controller display indicates that the hot-air sterilization was aborted. The

interior may still be hot. Do not touch the glass door and inner surfaces!

Fixed value

08.03.2013 05:05:06

Initial view after aborting the hot-air sterilization

(sample values).
For further information, press the “Info” button.

Setpoint

Actual value

Temperature [°C]

37.0 60.2

CO2 [Vol.-%]

5.0 5.0

Humidity [%rH]

85.5

User

Info

Menu

Sterilization

information

08.03.2013 05:05:06

Information menu “Sterilization information”.
Confirm the information with “Ok”.

Sterilization information

STERILIZATION ABORTED!

Close

Ok

Home


If the sterilization procedure is aborted prematurely, whether effective sterilization has occurred depends

on the time that has elapsed:

Aborting sterilization after less than 4 hours: Prevents effective sterilization.

Aborting sterilization after more than 4 hours: The unit is definitely in the cooling-down phase, meaning

that the necessary duration for the proper sterilization phase has occurred.


Aborting sterilization during the cooling-down phase (after more than 4 hours)
The duration of the entire sterilization is approx. 10 hours. If you want to shorten the sterilization

procedure in order to save time, you can abort it during the cooling-down phase, i.e. no sooner than after

4 hours. At this point, the inner temperature is still approx. 140 °C / 284°F.

CAUTION

Glass door and inner chamber become hot during sterilization.
Danger of burning.

Do NOT touch the glass door and inner surfaces for approx. 4 hours after aborting

sterilization.