6.4 Humidity control of the Permadry™ system

The Permadry™ system with its 2-pan water system guarantees a maximum humidity of up to 95 % r.H.

in the inner chamber, which remains condensation-free. This performance assumes an average ambient

temperature of 22 ±3 °C / 71.6 ±5.4 °F and a working temperature in the inner chamber of 37 °C /

98.6 °F.
You can increase or decrease humidity slightly if required.
To configure the humidity control of the Permadry™ system, go to

Menu > Humidity control

Fixed value

08.03.2013 05:05:06

Submenu “Humidity control”.
The current setting is displayed.
To change the setting, press the operating button.

Factory default setting: “Medium”.

..\ Humidity control Low/Medium/High

Humidity control Medium

Close

Home

Fixed value

08.03.2013 05:05:06

Submenu “Humidity control”.
Select the desired setting

and press the operating button

..\ Humidity control

Low

Medium

High

Close

Home

If the ambient temperature deviates by more than +/- 5 °C from the values recommended by

the manufacturer, the conditions for maximum air humidity with condensation-free inner

chamber are no longer guaranteed. Contact BINDER Service for assistance.