BINDER CB 60 User Manual
Page 112

CB (E6) 09/2013
Page 112/144
Fixed value
08.03.2013 05:05:06
Submenu “Start hot-air sterilization”.
This is a security question. Make sure that all
samples were removed from the CB before starting
the hot-air sterilization!
Then select “Really start (samples removed?)”
and press the operating button.
..\ Start hot-air sterilization
Do not start
Really start (samples removed?)
Close
Home
The controller returns to the initial view, and hot-air sterilization begins.
This symbol on the controller display indicates that the hot-air sterilization is running.
Sterilization
08.03.2013 05:05:06
Initial view during the heating-up phase of the hot-air
sterilization (sample values).
For further information, press the “Info” button.
Setpoint Actual value
Temperature [°C]
187.0 39.2
Section 01 End: 08.03.2013 15.48:19
CO2 [Vol.-%]
---- ----
Humidity [%rH]
---- ----
User
Info
Menu
Sterilization
08.03.2013 05:05:06
Information menu “Sterilization information”.
Confirm the information with “Ok”.
Sterilization information
DO NOT OPEN THE DOOR!
Close
Ok
Home