CB (E6) 09/2013

Page 124/144


Note: The recovery times of the gas concentrations inside the chamber following door opening, which are

indicated in the technical data (chap. 22.4), refer to a connection pressure of 2.0 bar / 29 psi. Decreasing

supply pressure down to the shift point (alarm point) of 0.3 bar / 4.4 psi results in longer recovery times.

Check the pressure displays of your gas supply. If very short recovery times are required or the door is

opened frequently, replace the gas cylinders promptly when the pressure decreases below 2.0 bar / 29

psi.
Fault description / indication Possible cause

Required measures

Low CO

2

outlet pressure

(< 0.3 bar)

CO

2

cylinder is not connected

correctly.

Confirm the alarm (chap. 11.4).

Correctly connect the gas cylinder.

Connected gas cylinder is

closed.

Confirm the alarm (chap. 11.4).

Open

the gas cylinder.

Connected gas cylinder is

empty.

Confirm the alarm (chap. 11.4).

Replace the gas cylinder. Observe

the safety instructions and the

maximum outlet pressure (chap. 4.3).

Alarm message

“Low pressure CO

2

Gas hose is dirty or obstructed.

Confirm the alarm (chap. 11.4).

Turn off the gas supply and remove

the gas connections. Check the tube

system for dirt accumulation or

obstruction, clean or replace it.

Controller malfunction.

Confirm the alarm (chap. 11.4). If

necessary, contact BINDER Service.

Pressure sensor system

defective.

Confirm the alarm (chap. 11.4).

Contact BINDER Service.

Low O

2

outlet pressure

(< 0.3 bar)

(unit with O

2

control)

O

2

cylinder is not connected

correctly.

Confirm the alarm (chap. 11.4).

Correctly connect the gas cylinder.

Connected gas cylinder is

closed.

Confirm the alarm (chap. 11.4).

Open

the gas cylinder.

Connected gas cylinder is

empty.

Confirm the alarm (chap. 11.4).

Replace the gas cylinder. Observe

the safety instructions and the

maximum outlet pressure (chap. 4.3).

Gas hose is dirty or obstructed.

Confirm the alarm (chap. 11.4).

Turn

off the gas supply and remove the

gas connections. Check the tube

system for dirt accumulation or

obstruction, clean or replace it.

Alarm message

“Low pressure O

2

Controller malfunction

Confirm the alarm (chap. 11.4).

If

necessary, contact BINDER Service.

Pressure sensor system

defective.

Confirm the alarm (chap. 11.4).

Contact BINDER Service.