Caution, Incubator – BINDER CB 60 User Manual

Page 108

background image

CB (E6) 09/2013

Page 108/144

Soapsuds may contain chlorides and must therefore NOT be used for cleaning.

With every decontamination method, always use adequate personal safety controls.

Following cleaning, leave the unit door open or remove the access port plugs.

The neutral cleaning agent may cause health problems in contact with skin and if ingested.

Follow the operating instructions and safety hints labeled on the bottle of the neutral

cleaning agent.

Recommended precautions: To protect the eyes use sealed protective goggles. Suitable protective

gloves with full contact: butyl or nitrile rubber, penetration time >480 minutes.

CAUTION

Contact with skin, ingestion.
Skin and eye damage due to chemical burns.

Do not ingest. Keep away from food and beverages.

Do NOT empty into drains.

Wear protective gloves and goggles.

Avoid skin contact.

Following use of the neutral cleaning agent and prior to hot-air sterilization, remove any agent

residues by using a moistened towel in order to avoid formation of permanent residues.


18.2 Decontamination / chemical disinfection of the CO

2

incubator

Disconnect the chamber from the power supply prior to decontamination. Disconnect the power plug.
You can use the following disinfectants:

Inner chamber

Standard commercial surface disinfectants free from acid or halides.
Alcohol based solutions.
We recommend using the disinfectant spray Art. No. 1002-0022.

For chemical disinfection, we recommend using the disinfectant spray Art. No. 1002-0022.
Any corrosive damage that may arise following use of other disinfectants is excluded from

liability by BINDER GmbH.

With every decontamination method, always use adequate personal safety controls.

This manual is related to the following products: