3 alarm status – BINDER CB 60 User Manual
Page 76

CB (E6) 09/2013
Page 76/144
11.3 Alarm status
An alarm message can appear in 3 different states:
“Set”
•
Active alarm.
•
The corresponding alarm icon is displayed in the initial view The buzzer sounds (if activated).
•
The “Info” button in the initial view leads to the “Alarms” submenu for alarm acknowledgement.
•
Press the “Reset” button in the “Alarms” submenu for alarm acknowledgement to mute the buzzer and
confirm the alarm.
“Acknowledged”
•
Active alarm.
•
The alarm was acknowledged. The buzzer is off.
•
The alarm cause is still valid. Therefore, the alarm icon remains displayed in the initial view.
•
The alarm message figures in the list of active alarms.
“Cleared”
•
The alarm cause has disappeared.
•
The alarm icon is not displayed any longer
•
The alarm message has disappeared from the list of active alarms.
•
The alarm message remains in the “Event list” for information.