3 storage, 4 location of installation and ambient conditions, Caution – BINDER CB 60 User Manual
Page 25

CB (E6) 09/2013
Page 25/144
CAUTION
Sliding or tilting of the unit.
Damage to the unit.
Risk of injury by lifting heavy loads.
Transport the unit in its original packaging only.
For moving or shipping, secure the unit with transport straps.
∅
Do NOT lift or transport the unit using the door handle, the door or the lower housing.
Lift the unit at the four lower corners with the aid of 4 people and place it on a rolling
pallet.
Move the unit to the desired location and lift it from the rolling pallet with the aid of four
people.
•
Permissible ambient temperature range during transport: 10 °C / 14 °F to +60 °C / 140 °F.
You can order transport packing for moving or shipping purposes from BINDER service.
3.3 Storage
Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 20.2).
•
Permissible ambient temperature range during storage: -10 °C / 14 °F to +60 °C / 140 °F.
•
Permissible ambient humidity: max. 70 % r.H., non-condensing
When after storage in a cold location you transfer the unit to its warmer installation site, condensation
may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature and
is completely dry.
3.4 Location of installation and ambient conditions
Notes on the location of installation
Set up the CO
2
incubator on a flat, even surface, free from vibration and in a well-ventilated, dry location.
The chambers are designed for setting up inside a building (indoor use).
Freestanding CB incubators are suitable for installation on tables or on the optionally available stand
(height 200 mm / 0.5 ft). Note: The site of installation must be capable of supporting the unit’s weight (see
technical data, chap.22.4).
Align the unit using a spirit level to ensure even covering of the cell-cultures with the medium. For this
purpose, manually adjust the four incubator feet.
CO
2
incubators can be stacked on top of each other (two units maximum). For safe stacking that is easy
to maintain, use the original BINDER stacking stand (chap. 15.8.1) or the stacking adapter (chap. 15.8.2).
CAUTION
Sliding of stacked unit.
Damage to the units.
When stacking, use rubber pads (Art. No. 8012-0376) for the feet of the upper
incubator.
To completely separate the unit from the power supply, you must disconnect the power plug. Install the
unit in a way that the power plug is easily accessible and can be easily pulled in case of danger.
In order to avoid contamination, never place the unit directly on the floor.