3 aborting the hot-air sterilization prematurely, Caution – BINDER CB 60 User Manual
Page 113

CB (E6) 09/2013
Page 113/144
Opening the outer door during sterilization leads to aborting the automatic sterilization cycle.
CAUTION
Interruption of the temperature reaction time.
Ineffective sterilization.
∅
Do NOT open the unit doors during sterilization.
CAUTION
Glass door and inner chamber become hot during sterilization.
Danger of burning.
∅
Do NOT touch the glass door and inner surfaces during sterilization
18.3.3 Aborting the hot-air sterilization prematurely
Three events lead to aborting the automatic sterilization cycle prematurely:
•
A manual abortion via the controller menu
•
Opening the outer door
•
Turning off the incubator at the main power switch, or a power failure
To manually abort the hot-air sterilization cycle, go to
Menu > Hot-air sterilization
or
Quick menu > Hot-air sterilization
Sterilization
08.03.2013 05:05:06
Submenu “Hot-air sterilization”.
To abort the hot-air sterilization, select “Stop”
and press the operating button.
..\ Hot-air sterilization
Start
Stop
Close
Home
Sterilization
08.03.2013 05:05:06
Submenu “Stop hot-air sterilization”.
This is a security question.
To abort the hot-air sterilization, select “Really stop”
and press the operating button.
..\ Stop hot-air sterilization
Do not stop
Really stop
Close