11.6 Zero-voltage relay alarm output

The CO

2

incubator is equipped with a zero-voltage relay output which permits the transmission of some

alarm messages to a central monitoring system.
The connection is realized as a DIN socket (9) on the unit rear. A suitable DIN plug is enclosed.

Figure 16: Pin configuration of the DIN socket (9) on the unit rear

ALARM

Pin 1: Pole, Pin 2: Break relay, Pin 3: Make contact

In case there is no alarm, contact 1 closes with contact 3.
Closing contact 1 with contact 2 switches the zero-voltage relay alarm output.
Maximum loading capacity of the switching contacts: 24V AC/DC – 2.5A

DANGER

Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.

Do NOT exceed the maximum switching load of 24V AC/DC – 2.5A.

Do NOT connect any devices with a higher loading capacity.

The zero-voltage relay alarm output switches at the following events:
Displayed icon

Error / event when icon is

displayed

Switching the alarm contact

Power failure

immediately

Tolerance range alarm of

temperature, CO

2

, O

2

(unit

with O

2

control) or humidity

after 10 min from error

Door open

for more than the

set alarm delay time (chap.

11.5.5, factory setting: 1

minute)

after 10 minutes from door opening

In case of a tolerance range alarm or a door open alarm, the alarm message on the controller display

remains on during the alarm transmission via the zero-voltage relay outputs.
As soon as the cause of the alarm is identified and resolved, you can reset the alarm transmission via the

zero-voltage relay outputs together with the alarm message on the controller display by hitting the

“RESET” key.
In case of a power failure, transmission of the alarm via zero-voltage relay outputs remains active for the

duration of the power failure. After power returns, contact 1 closes automatically with contact 3.

When using the communication software APT-COM™ 3 DataControlSystem (option, chap.

15.1) via the Ethernet interface or optional RS 422 interface of the CO

2

incubator for data

acquisition, only the alarm message is recorded in the APT-COM™ protocol.

Set the tolerance limits for limit alarms by APT-COM™ 3 separately in the APT-COM™ 3

measuring window.