4 sterilization, 5 humidity – BINDER CB 60 User Manual
Page 127

CB (E6) 09/2013
Page 127/144
21.4 Sterilization
Fault description / indication Possible cause
Required measures
The sterilization is running.
Sterilization cycle is in the
heating up or holding phase
Confirm the message (chap. 11.4).
Do not open the door.
Notification
“DO NOT OPEN THE
DOOR !”.
The sterilization is successfully
completed: Interior is sterilized.
Heating up and holding phase
of the sterilization cycle
completed.
Confirm the message (chap. 11.4).
The inner chamber and parts
inside can still be hot. Do not
touch.
You can open the doors. Let the unit
cool down (chap. 18.3) and take it
into operation again (chap. 5).
Notification
“STERILIZATION
FINISHED”.
Abortion of sterilization. Interior
is not sterilized.
Manual abortion der of the hot-
air sterilization.
Confirm the message (chap. 11.4).
The inner chamber and parts
inside can still be hot. Do not
touch.
Interior is not sterilized. Start again
the hot-air sterilization or disinfect the
interior (chap. 18.2)
Notification
“STERILIZATION
ABORTED”.
Outer door opened during hot-
air sterilization.
Confirm the message (chap. 11.4).
Close the unit door and start again
the hot-air sterilization.
21.5 Humidity
Fault description / indication Possible cause
Required measures
No or too low humidity inside.
Permadry™ pan empty.
Fill the outer Permadry™ pan with
water up to the marking on the inner
pan with distilled, sterile water. The
pan must have thorough contact to
the bottom of the inner chamber.
Chap. 4.2.
Humidity set too low.
Increase humidity (chap. 6.4)
The displayed actual humidity
value deviates largely
compared with a reference
method.
Humidity sensor not adjusted.
Calibrate and adjust humidity sensor.
Humidity sensor defective.
Contact BINDER Service.
Condensations inside the
chamber.
Permadry™ pan filled with
water when incubator is not
operating.
Empty Permadry™ pan when
incubator is not operating.
Door gaskets defective.
Replace door gaskets
Doors not closed properly.
Close door properly.
Humidity set too high.
Reduce humidity (chap. 6.4)
Condensation on the door.
Unit placed on very cold floor.
Place the unit on a BINDER stand to
increase the distance to the floor.
Door gaskets defective.
Replace door gaskets
Doors not closed properly.
Close door properly.
Humidity set too high.
Reduce humidity (chap. 6.4)
Door heating set too low.
Contact BINDER Service to adjust
the door heating