beautypg.com

Troubleshooting, 1 heating – BINDER CB 60 User Manual

Page 122

background image

CB (E6) 09/2013

Page 122/144

21. Troubleshooting

Only qualified service personnel authorized by BINDER must perform repair.

Repaired units must comply with the BINDER quality standards.


21.1 Heating

Fault description / indication Possible cause

Required measures

Excess temperature. Having

reached the setpoint, the

temperature exceeds the

setpoint by more than the set

tolerance range value (more

than 10 min.).

Site of installation too warm.

Difference between the set

temperature and the ambient

temperature too low.

Confirm the alarm (chap. 11.4).

Difference between the set

temperature and the ambient

temperature must be at least 7 °C /

12.6 °F.

Alarm message

“Temp. range”

Excess temperature.

Chamber heating permanently,

exceeding the setpoint.

Temporary disturbance of the

temperature control.

Confirm the alarm (chap. 11.4). Turn

off the chamber. Open the chamber

doors for approx.. 5 minutes and turn

on the chamber again.

Confirm the

alarm (chap. 11.4). Upon renewed

alarm, contact BINDER service.

Notification

“Heating active”.

Alarm message

“Temp. range”

Controller defective.

Confirm the alarm (chap. 11.4).

Contact BINDER service.

Semiconductor relay defective
Temperature controller not

adjusted.

Confirm the alarm (chap. 11.4).

Calibrate and adjust the temperature

controller.

Excess temperature.

Overtemperature safety

controller class 3.1 responds.

Safety controller setpoint value

exceeded.

Confirm the alarm (chap. 11.4).

Check setting of temperature setpoint

and of the safety controller class 3.1

setpoint. If appropriate, select

suitable value.

Alarm message

“Safety controller”

Too much external heat load.

Confirm the alarm (chap. 11.4).

Reduce heat load.

Controller defective.

Confirm the alarm (chap. 11.4).

Contact BINDER service.

Semiconductor relay defective
Safety controller defective.

Too low temperature.

Setpoint temperature is not

reached after specified time.

Unit door not properly closed.

Completely close unit door.

Semiconductor relay defective. Contact BINDER service

Notification

“Heating active”.

Door gasket defective.

Replace door gasket.

Temperature controller not

adjusted.

Calibrate and adjust the temperature

controller.

This manual is related to the following products: