To activate or deactivate the O
2
control and O
2
/ N
2
pressure alarms, go to
Menu > Controller mode > O2 control
Fixed value
08.03.2013 05:05:06
Submenu “O2 control”.
The current settings are displayed.
To change the settings, press the operating button.
The modified settings are displayed
“O2 control: On” = O
2
control is active.
“O2 control: Off” = O
2
control is deactivated.
“O2 pressure alarm: On” or “N2 pressure alarm: On”
= the corresponding pressure alarm is activated.
“O2 pressure alarm: Off” or “N2 pressure alarm: Off”
= the corresponding pressure alarm is deactivated.
..\ O2 control
O2 control: On
O2 pressure alarm: On
N2 pressure alarm: On
Close
Home
Go back to the initial view with “Home”.
Fixed value
08.03.2013 05:05:06
Initial view with deactivated O
2
control
(sample values).
Also the pressure alarms for O
2
and N
2
are
deactivated.
Setpoint
Actual value
Temperature [°C]
37.0 37.2
CO2 [Vol.-%]
5.0
5.0
Humidity [%rH]
85.5
O2 [Vol.-%]
20.7
User
Quick menu
Menu
To enter the O
2
setpoint, first activate the O
2
control again.
With active O
2
control, the pressure alarms for O
2
and N
2
can be deactivated individually.
Fixed value
08.03.2013 05:05:06
Initial view with activated O
2
control and deactivated
O
2
pressure alarm (sample values).
Setpoint
Actual value
Temperature [°C]
37.0 37.2
CO2 [Vol.-%]
5.0
5.0
Humidity [%rH]
85.5
O2 [Vol.-%]
25.0 25.0
User
Quick menu
Menu