CB (E6) 09/2013

Page 109/144


In case of contamination of the interior by biologically or chemically hazardous material, there are two

possible procedures depending on the type of contamination and charging material.
(1) Spray the inner chamber with an appropriate disinfectant.

Before start-up, the unit must be absolutely dry and ventilated, as explosive gases may form during

the decontamination process.

(2) If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning, or

have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave.

In case of eye contact, the disinfectant spray may cause eye damage due to chemical

burns. Follow the operating instructions and safety hints labeled on the bottle of the

disinfectant spray.

Recommended precautions: To protect the eyes use sealed protective goggles.

CAUTION

Eye contact.
Eye damage due to chemical burns.

Do NOT empty into drains.

Wear protective goggles.

After using the disinfectant spray, allow the incubator to dry thoroughly, and aerate it

sufficiently.

Following frequent use of the disinfectant spray and prior to hot-air sterilization, remove any

agent remainder by using the neutral cleaning agent and then a moistened towel to avoid

formation of permanent residues.