3 gas concentration – BINDER CB 60 User Manual

Page 125

background image

CB (E6) 09/2013

Page 125/144


Fault description / indication Possible cause

Required measures

Low N

2

outlet pressure

(< 0.3 bar)

(unit with O

2

control)

N

2

cylinder is not connected

correctly.

Confirm the alarm (chap. 11.4).

Correctly connect the gas cylinder.

Connected gas cylinder is

closed.

Confirm the alarm (chap. 11.4).

Open

the gas cylinder.

Connected gas cylinder is

empty.

Confirm the alarm (chap. 11.4).

Replace the gas cylinder. Observe

the safety instructions and the

maximum outlet pressure (chap. 4.3).

Gas hose is dirty or obstructed.

Confirm the alarm (chap. 11.4).

Turn

off the gas supply and remove the

gas connections. Check the tube

system for dirt accumulation or

obstruction, clean or replace it.

Alarm message

“Low pressure N

2

Controller malfunction

Confirm the alarm (chap. 11.4).

If

necessary, contact BINDER Service.

Pressure sensor system

defective.

Confirm the alarm (chap. 11.4).

Contact BINDER Service.


21.3 Gas concentration

Fault description / indication Possible cause

Required measures

Excess CO

2

concentration.

Having reached the setpoint,

CO

2

exceeds the setpoint by

more than the set tolerance

range value (more than 10

min.)

Temporary disturbance of the

CO

2

control.

Confirm the alarm (chap. 11.4). Turn

off the chamber. Open the chamber

doors for approx. 5 minutes. Observe

the general information for safe

handling of CO

2

(chap. 1.6). Turn on

the chamber again.

Upon renewed

alarm, contact BINDER service.

Alarm message

“CO

2

range”

CO

2

sensor defective.

Confirm the alarm (chap. 11.4).

Contact BINDER Service.

Excess O

2

concentration.

Having reached the setpoint,

O

2

exceeds the setpoint by

more than the set tolerance

range value (more than 10

min.) (unit with O

2

control)

Temporary disturbance of the

O

2

control.

Confirm the alarm (chap. 11.4). Turn

off the chamber. Open the chamber

doors for approx. 5 minutes. Observe

the general information for safe

handling of oxygen (chap. 1.6)..

Prevent oxygen enrichment in the

ambiance of the unit. Turn on the

chamber again.

Upon renewed alarm,

contact BINDER service.

Alarm message

“O

2

range”

O

2

sensor defective.

Confirm the alarm (chap. 11.4).

Contact BINDER Service.

Recovery time (up to 5 vol.-%

CO

2

) after doors were open for

2 minutes is < 2 minutes

CO

2

sensor system defective.

Contact BINDER Service.