15. Options
15.1 Communication software APT-COM™ 3 DataControlSystem (option)
The unit is regularly equipped with an Ethernet interface (6) that can connect the BINDER communication
software APT-COM™ 3 DataControlSystem. The MAC Address is indicated below the Ethernet interface.
The actual temperature, and gas concentration values are given in adjustable intervals. Up to 30
chambers with RS 422 interface can be cross linked. For further information, refer to the operating
manual of the BINDER communication software APT-COM™ 3.
15.2 RS 422 interface (option)
With this option, the chamber is equipped with a serial interface RS 422 (5), that can connect the BINDER
communication software APT-COM™ 3 DataControlSystem. The MAC Address is indicated below the
Ethernet interface. The actual temperature, and gas concentration values are given in adjustable
intervals. For further information, refer to the operating manual of the BINDER communication software
APT-COM™ 3.
The connection to a computer is established using the CB interface via an interface converter.
Pin allocation of the RS 422 interface (5) on the
unit rear
Pin 2:
RxD (+)
Pin 3:
TxD (+)
Pin 4:
RxD (-)
Pin 5:
TxD (-)
Pin 7:
Ground
15.3 Silicone access ports 30 mm / 1.18 in, closable from both sides with silicon
plugs (option)
(a)
(b)
(c)
Figure 17: Positions of the optional silicon access ports left (a), rear (b), and right (c)
A warning sticker is located above each access port.