3 checking the humidity system fan, 4 gas inlet fine filter, 5 sending the unit back to binder gmbh – BINDER CB 60 User Manual
Page 117
CB (E6) 09/2013
Page 117/144
19.3 Checking the humidity system fan
The user should regularly perform the following checks:
Hold a sheet of paper in front of the ventilation silts at the bottom rear of the unit. If the sheet of paper is
suctioned inwards, the fan operates correctly. When opening the door, the fan automatically turns off.
19.4 Gas inlet fine filter
When the unit is in operation, the incoming gas passes through a fine gas filter (aseptic filter, filtration
efficiency 99.99 %, particle size 0.45 µm).This fine filter prevents dirt accumulating in the gas inlet valves
and the tubes leading into the inner chamber, which could be in the gas cylinder or in the supply tubes.
When using gas with a technical grade of 99.5 %, we recommend changing the fine gas filter once a year.
Please consult BINDER Service. When using gases with less pureness, the changing intervals should be
shorter.
19.5 Sending the unit back to BINDER GmbH
If you return a BINDER product to us for repair or any other reason, we will only accept the product upon
presentation of an authorization number that has previously been issued to you. An authorization number
will be issued after receiving your complaint either in writing or by telephone
prior to your sending the
BINDER product back to us. The authorization number will be issued following receipt of the information
below:
•
BINDER product type and serial number
•
Date of purchase
•
Name and address of the dealer from which you bought the BINDER product
•
Exact description of the defect or fault
•
Complete address, contact person and availability of that person
•
Exact location of the BINDER product in your facility
•
A contamination clearance certificate (chap. 23) must be faxed in advance
The authorization number must be applied to the packaging in such a way that it can be easily recognized
or be recorded clearly in the delivery documents.
For security reasons we cannot accept a unit delivery if it does not carry an authorization
number.