Maintenance and service, 1 maintenance intervals, service, 2 checking the air jacket heating fan – BINDER CB 60 User Manual
Page 116: Danger

CB (E6) 09/2013
Page 116/144
Therefore, without using the stacking stand or the stacking adapter, safe incubation is not possible during
sterilization of the other unit. We recommend using a stacking stand (chap. 15.8.1) or a stacking adapter
(chap. 15.8.2).
19. Maintenance and service
19.1 Maintenance intervals, service
DANGER
Electrical hazard.
Danger of death.
∅
The unit must NOT become wet during operation or maintenance works.
∅
Do NOT remove the rear panel of the unit.
Before conducting maintenance work, turn off the unit at the main power switch and
disconnect the power plug.
Any maintenance work must be conducted by licensed electricians or experts
authorized by BINDER.
Ensure regular maintenance work is performed at least once a year.
When the unit requires maintenance a corresponding note is displayed on the controller screen. After
confirming with “OK”, this message will come up daily until reset by the service technician.
The warranty becomes void if maintenance work is conducted by non-authorized personnel.
Replace the door gasket only when cold. Otherwise, the door gasket may become damaged.
With an increased amount of dust in the ambient air, clean the Peltier fan grid (7) by suction or blowing
several times a year.
We recommend taking out a maintenance agreement. Please consult BINDER Service.
BINDER telephone hotline:
+49 (0) 7462 2005 555
BINDER fax hotline:
+49 (0) 7462 2005 93555
BINDER e-mail hotline:
BINDER service hotline USA:
+1 866 885 9794 or +1 631 224 4340 x3 (toll-free in the USA)
BINDER service hotline Asia Pacific:
+852 39070500 or +852 39070503
BINDER service hotline Russia and CIS +7 495 98815 17
BINDER Internet website
http://www.binder-world.com
BINDER address
BINDER GmbH, post office box 102, D-78502 Tuttlingen
International customers, please contact your local BINDER distributor.
The gas sensors are especially adjusted for the specific incubator. When exchanging a
sensor, you must repeat control adjustment for CO
2
and O
2
(unit with O
2
control).
19.2 Checking the air jacket heating fan
The user should regularly perform the following checks:
When the unit is operating, you will be able to observe the air jacket heating fan
centered at the top of the
unit by looking through the ventilation slides at the rear. During unit operation, the fan must turn counter-
clockwise continuously. For better monitoring, you can turn off the unit and wait until the fan has stopped.