Fault description / indication Possible cause

Required measures

Condensation on the divided

inner glass door, gas-proof.

Unit placed on very cold floor.

Place the unit on a BINDER stand to

increase the distance to the floor.

Door gaskets defective.

Replace door gaskets

Doors not closed properly.

Close door properly.

Humidity set too high.

Reduce humidity (chap. 6.4).

Door heating set too low.

Contact BINDER Service to adjust

the door heating

Excess Humidity. After the

humidity was situated within the

defined tolerance range, it

deviates longer than 10 min.

from the tolerance range.

Temporary disturbance of the

humidity control.

Confirm the alarm (chap. 11.4).

Turn

off the chamber and turn it on again.

Upon renewed alarm, contact

BINDER service.

Controller defective.

Confirm the alarm (chap. 11.4).

Contact BINDER service.

Alarm message:

„Humid range“

Humidity sensor not adjusted.

Confirm the alarm (chap. 11.4).

Calibrate and adjust humidity sensor.

Humidity sensor defective

Confirm the alarm (chap. 11.4).

Contact BINDER service.

Too low humidity. After the

humidity was situated within the

defined tolerance range, it

deviates longer than 10 min.

from the tolerance range

or the

humidity doesn’t reach the

tolerance range within 3 hours

from turning on the unit or

closing the door.

Unit door not properly closed.

Confirm the alarm (chap. 11.4).

Completely close unit door.

Door gasket defective.

Confirm the alarm (chap. 11.4).

Replace door gasket.

Permadry™ pan empty.

Confirm the alarm (chap. 11.4).

Fill

the outer Permadry™ pan with water

up to the marking on the inner pan

with distilled, sterile water. The pan

must have thorough contact to the

bottom of the inner chamber, see

chap. 4.2.

Temporary disturbance of the

humidity control.

Confirm the alarm (chap. 11.4).

Turn

off the chamber and turn it on again.

Upon renewed alarm, contact

BINDER service.

Controller defective.

Confirm the alarm (chap. 11.4).

Contact BINDER service.

Alarm message:

„Humid range“

Humidity sensor not adjusted.

Confirm the alarm (chap. 11.4).

Calibrate and adjust humidity sensor.

Humidity sensor defective.

Confirm the alarm (chap. 11.4).

Contact BINDER service.