Fault description / indication Possible cause
Required measures
Condensation on the divided
inner glass door, gas-proof.
Unit placed on very cold floor.
Place the unit on a BINDER stand to
increase the distance to the floor.
Door gaskets defective.
Replace door gaskets
Doors not closed properly.
Close door properly.
Humidity set too high.
Reduce humidity (chap. 6.4).
Door heating set too low.
Contact BINDER Service to adjust
the door heating
Excess Humidity. After the
humidity was situated within the
defined tolerance range, it
deviates longer than 10 min.
from the tolerance range.
Temporary disturbance of the
humidity control.
Confirm the alarm (chap. 11.4).
Turn
off the chamber and turn it on again.
Upon renewed alarm, contact
BINDER service.
Controller defective.
Confirm the alarm (chap. 11.4).
Contact BINDER service.
Alarm message:
„Humid range“
Humidity sensor not adjusted.
Confirm the alarm (chap. 11.4).
Calibrate and adjust humidity sensor.
Humidity sensor defective
Confirm the alarm (chap. 11.4).
Contact BINDER service.
Too low humidity. After the
humidity was situated within the
defined tolerance range, it
deviates longer than 10 min.
from the tolerance range
or the
humidity doesn’t reach the
tolerance range within 3 hours
from turning on the unit or
closing the door.
Unit door not properly closed.
Confirm the alarm (chap. 11.4).
Completely close unit door.
Door gasket defective.
Confirm the alarm (chap. 11.4).
Replace door gasket.
Permadry™ pan empty.
Confirm the alarm (chap. 11.4).
Fill
the outer Permadry™ pan with water
up to the marking on the inner pan
with distilled, sterile water. The pan
must have thorough contact to the
bottom of the inner chamber, see
chap. 4.2.
Temporary disturbance of the
humidity control.
Confirm the alarm (chap. 11.4).
Turn
off the chamber and turn it on again.
Upon renewed alarm, contact
BINDER service.
Controller defective.
Confirm the alarm (chap. 11.4).
Contact BINDER service.
Alarm message:
„Humid range“
Humidity sensor not adjusted.
Confirm the alarm (chap. 11.4).
Calibrate and adjust humidity sensor.
Humidity sensor defective.
Confirm the alarm (chap. 11.4).
Contact BINDER service.