11. Notifications and Alarms
11.1 Notifications overview
Icon
Signification
Icon
Signification
Fixed value operation
O
2
/N
2
control deactivated. No O
2
concentration alarms will occur.
(unit with O
2
control)
Heating active
CO
2
pressure alarm deactivated
Key lock activated
O
2
pressure alarm deactivated
(unit with O
2
control)
Copying data via USB
N
2
pressure alarm deactivated
(unit with O
2
control)
Interior socket turned on (with option interior
socket)
Icon
Notifying message
Signification
DO NOT OPEN
THE DOOR !
Hot air sterilization running
STERILIZATION
FINISHED
Hot air sterilization successful
The inner chamber and parts inside can still be hot. Do not touch.
STERILIZATION
ABORTED
Hot air sterilization cancelled
The inner chamber and parts inside can still be hot. Do not touch.
11.2 Alarm messages overview
Icon
Alarm message
Signification
Safety controller
overtemperature
Overtemperature safety controller alarm (class 3.1): selected value of the
safety controller exceeded.
Temp. range
Having reached the setpoint, the temperature deviates by more than the
set tolerance range value (chap. 11.5.6) and longer than 10 min. from the
setpoint
or the temperature doesn’t reach the tolerance range within 3
hours from turning on the unit or closing the door
CO2 range
Having reached the setpoint, the CO
2
concentration deviates by more than
the set tolerance range value (chap. 11.5.6) and longer than 10 min. from
the setpoint
or the CO
2
concentration doesn’t reach the tolerance range
within 3 hours from turning on the unit or closing the door
O2 range
Having reached the setpoint, the O
2
concentration deviates by more than
the set tolerance range value (chap. 11.5.6) and longer than 10 min. from
the setpoint
or the O
2
concentration doesn’t reach the tolerance range
within 3 hours from turning on the unit or closing the door (unit with O
2
control)