Remove the rubber cover (b) by pulling it off.

Unit rear

Figure 12: Connecting the hose lead to the gas cylinder

Now fit the hose nozzle (c) in the quick acting closure socket. To remove the connection, pull the hose

nozzle off the quick acting closure socket.

4.3.5 Gas cylinder connection kits (option)

Gas cylinder connection kits are available for CO

2

(Art. No. 8012-0014), O

2

(Art. No. 8012-0015) and N

2

(Art. No. 8012-0016).

The connection kit includes the following parts for connecting a gas cylinder

to the CO

2

incubator:

Pressure reducer with manometers for cylinder pressure (high pressure

gauge) and outlet pressure (low pressure gauge)

5 m pressure hose with pre-assembled hose nozzle for quick acting

closure socket

1 hose clamp to connect the gas hose to the pressure reducer



Figure 13: Gas cylinder connection kit

Instructions 7001-0249 included with the connection kit describe connecting and setting the

gas cylinder connection kit.

CAUTION

Excessive outlet pressure > 2.5 bar / 36 psi.
Damage to the unit.

The outlet pressure must NOT exceed the indicated value of 2.5 bar / 36 psi.

Before connecting, check the outlet pressure on the pressure reducer of the cylinder.

Adjust the outlet pressure to 2.0 bar / 29 psi above the ambient pressure.

The gas connections must be established by qualified personnel who are trained in handling

the respective gases and familiar with the required safety measures.