Fault description / indication Possible cause

Required measures

Too low CO

2

concentration.

Having reached the setpoint,

CO

2

falls below the setpoint by

more than the set tolerance

range value (more than 10

min.)

or CO

2

doesn’t reach the

tolerance range within 3 hours

from turning on the unit or

closing the door.

Temporary disturbance of the

CO

2

control.

Confirm the alarm (chap. 11.4).

Turn

off the chamber. After approx. 5

minutes turn on the chamber again.

Upon renewed alarm, contact

BINDER service.

Alarm message

“CO

2

range”

CO

2

sensor defective.

Confirm the alarm (chap. 11.4).

Contact BINDER Service.

Too low O

2

concentration.

Having reached the setpoint,

O

2

falls below the setpoint by

more than the set tolerance

range value (more than 10

min.)

or O

2

doesn’t reach the

tolerance range within 3 hours

from turning on the unit or

closing the door (unit with O

2

control)

Temporary disturbance of the

O

2

control.

Confirm the alarm (chap. 11.4).

Turn

off the chamber. After approx. 5

minutes turn on the chamber again.

Upon renewed alarm, contact

BINDER service.

Alarm message

“O

2

range”

O

2

sensor defective.

Confirm the alarm (chap. 11.4).

Contact BINDER Service.

Too low gas concentration.

The concentration of CO

2

or O

2

(unit with O

2

control) does not

reach the adjusted set value.

Door gaskets defective.

Replace door gaskets

Doors not closed properly.

Close door properly.

Connected gas cylinder is

empty or not opened.

Open or replace gas cylinder.

Gas cylinder is not connected

correctly.

Correctly connect the gas cylinder.

Gas hose is dirty or obstructed.

Check the tube system for dirt

accumulation or obstruction, clean or

replace it.

Recovery time (up to 5 vol.-%

CO

2

) after doors were open for

2 minutes is > 10 minutes.

Obstructed gas supply.

Check gas supply (cylinder,

connections, hose system).

Insufficient CO

2

input pressure

Unusually high gas

consumption.

Door gaskets defective.

Replace door gaskets

Gas sensor not adjusted.

Calibrate the sensor.

Gas sensor defective.

Contact BINDER Service

Gas fine filter not connected

correctly.

The displayed actual value of

CO

2

deviates largely compared

with a reference method.. The

pH indicator of the cell medium

changes its normal color.

CO

2

control not adjusted.

Calibrate and adjust CO

2

control.

CO

2

sensor defective.

Transfer the cultures to another

incubator and contact BINDER

Service.

The displayed actual value of

O

2

(unit with O

2

control)

deviates largely compared with

a reference method.

O

2

control not adjusted.

Calibrate and adjust O

2

control.

O

2

sensor defective.

Transfer the cultures to another

incubator and contact BINDER

Service.