14.2.2 Setting the safety controller

To display and to change the current safety controller settings in the “safety controller” submenu, go to

Menu > Safety controller

Safety controller mode: selection between Limit (absolute) and Offset (relative)

Fixed value

08.03.2013 05:05:06

Submenu “Safety controller”
Select “Mode”

and press the operating button.

..\ Safety controller

Mode

Setpoint

Show settings

Close

Home

Fixed value

08.03.2013 05:05:06

Submenu “Safety controller mode”.
The current safety controller mode is displayed:

“Limit (absolute)” or “Offset (relative)”
To change the mode, press the operating button.
The modified safety controller mode is displayed.

..\ Safety controller mode

Limit (absolute)

Close

Home

When changing the safety controller mode, the safety controller setpoint which had been active

before in this mode becomes active again.


Entering the safety controller setpoint

Fixed value

08.03.2013 05:05:06

Submenu “Safety controller”
Select “Setpoint”

and press the operating button.

..\ Safety controller
Mode

Setpoint

Show settings

Close

Home


You can also access this submenu to directly enter the safety controller setpoint via

Quick menu > Safety controller setpoint