Warning, Caution – BINDER CB 60 User Manual
Page 97

CB (E6) 09/2013
Page 97/144
Heat emission of electrical devices connected inside the CO
2
incubator may modify the
temperature range.
The maximum load must not exceed 3 Amp.
WARNING
Overload of contacts.
Damage to contacts and connection socket.
Electrical hazard.
Danger of death.
∅
Do NOT exceed the maximum load of 3 A.
∅
Do NOT connect any devices with a nominal current > 3 A.
Turning on and off the voltage of the interior socket is possible at the chamber controller through
Menu >
Interior socket (optional)
, see chap. 6.5.
Turning off the incubator at the main power switch also switches the interior socket voltage-free.
Figure 19: Interior socket (G) (front view)
Figure 20: Supplied plug (front view)
Electrical data socket and plug: IP system of protection 65, 230 V 1N ~ 50-60 Hz
CAUTION
Risk of short circuit.
Damage to the unit.
Use delivered plug only (IP protection type 65).
If the socket is not used, close it with the waterproof lid.
When inserting a heat load into the inner chamber by introducing electrical devices such as shakers or
rollers, the ambient temperature of the incubator must not exceed 25 °C / 77°F. Otherwise, temperature
control to 37 °C / 98.6°F cannot be assured. When operating electrical devices inside the incubator,
always check the desired incubation temperature
before introducing any cells.
CAUTION
Excessive inner temperature > 37 °C / 98.6°F when operating electrical devices
inside the incubator.
Damage to the cell cultures.
When operating electrical devices inside the incubator: Ambient temperature # 25 °C /
77°F.
Check the temperature observation before introducing any cells.
Pin 2 (L)
Pin 1 (N)
Pin 3 (PE)
Pin 2 (L)
Pin 1 (N)
Pin 3 (PE)