7.4 Entering the setpoints via general menu

To enter setpoints via general menu, go to

Menu > Setpoints

Fixed value

08.03.2013 05:05:06

Submenu “Setpoints”.
(item “O2 setpoint” only coming with unit with O

2

control)
Select the desired parameter

and press the operating button.

..\ Setpoints

Temperature setpoint

CO2 setpoint

O2 setpoint

Close

Home

When trying to enter a setpoint in the „Controller off“ operating mode, a notification window

shows “Controller mode is OFF!”. Press the operating button to confirm with “Ok” and change

the operating mode to “Fixed value” (chap. 6.2.1).



Temperature setting
To enter the temperature setpoint, go to

Menu > Setpoints > Temperature setpoint

Fixed value

08.03.2013 05:05:06

Entry menu “Temperature setpoint”.
Select each number with the operating button and

press the operating button to confirm.
Setting range: 20 °C / 68 °F up to 60 °C / 140 °F.
Press the “Ok” button to confirm.

..\ Temperature setpoint

3

7.0

[°C]

3

End Ok 0 1 2

4 5 6 7 8 9 Ins

Close

Ok

Home

When entering a value outside the setting range, the message “invalid value” appears. Press the

operating button to confirm with “Ok” and repeat the entry with a correct value.