Ole_link1, Ole_link2 – BINDER CB 60 User Manual
Page 132

CB (E6) 09/2013
Page 132/144
Unit size
60
160
220
Humidity data
Humidity
% r.H.
90 to 95
90 to 95
90 to 95
O
2
Data
O
2
range
vol.-% O
2
0.2 to 95
0.2 to 95
0.2 to 95
Setting accuracy with inlet pressure 2 bar
vol.-% O
2
0.1
0.1
0.1
Recovery time after door was opened for
30 sec at 1 vol-% O
2
1)
minutes
8
12
18
Recovery time after door was opened for
30 sec at 5 vol-% O
2
1)
minutes
8
12
18
O
2
measurement
ZrO
2
ZrO
2
ZrO
2
Connection hose nozzle DN6 for O
2
/N
2
connection to the unit for hose with internal
diameter
mm / inch
6 / 0.24
6 / 0.24
6 / 0.24
Electrical data CB
IP system of protection acc. to EN 60529
IP
20
20
20
Nominal voltage (±10 %)
V
200-240 1N~ 200-240 1N~ 200-240 1N~
Power frequency
Hz
50/60
50/60
50/60
Nominal power
kW
1.00
1.30
1.50
Energy consumption 2) at 37 °C/ 98.6 °F
Wh/h
80
100
120
Power plug (IEC connector plug)
Shock-proof plug
Installation category acc. to IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Unit fuse 4x20mm / semi time-lag (M)
Amp
10
10
10
Electrical data CB-UL (for the USA and Canada)
Nominal voltage (±10 %) 60 Hz / 1N
V
100-120 1N~ 100-120 1N~ 100-120 1N~
Power frequency
Hz
50/60
50/60
50/60
Power plug
NEMA
5-15P
5-20P
5-20P
Nominal power
kW
1.00
1.30
1.50
Installation category acc. to IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Unit fuse 6.3 X 32 mm / 250V / super-time-
lag TT
Amp
16 external
16 external
16 external
Additional temperature fuse class 1 (DIN 12880)
internal
internal
internal
Legend:
1) The recovery times of the gas concentrations inside the chamber after the door is opened coincide
with a connection pressure of 2.0 bar / 29 psi. Decreasing supply pressure results in longer recovery
times.
2) Use this value for sizing air condition systems.
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
22 ±3 °C / 71.6 ±5.4 °F and a power supply voltage fluctuation of ±10. The temperature data is
determined in accordance to BINDER factory standard and DIN 12880, observing the recommended wall
clearances of 10 % of the height, width and depth of the inner chamber.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.