4 electrical connection, Caution – BINDER CB 60 User Manual
Page 36
CB (E6) 09/2013
Page 36/144
4.4 Electrical connection
The CO
2
incubator comes with an IEC connector plug.
Model
Art. No.
(x = 0 or 1)
Power plug
Voltage
+/-10 %
Power
frequency Unit fuse
CB 60
9x40-0088
Shock-proof plug
230 V (1N~)
50/60 Hz
16 A
CB 60-UL
9x40-0089
NEMA 5-15P
100-120 V (1N~) 60 Hz
16 A
CB 60
with O
2
control
9x40-0090
Shock-proof plug
230 V (1N~)
50/60 Hz
16 A
CB 60-UL
with O
2
control
9x40-0091
NEMA 5-15P
100-120 V (1N~) 60 Hz
16 A
CB 160
9x40-0092
Shock-proof plug
230 V (1N~)
50/60 Hz
16 A
CB 160-UL
9x40-0093
NEMA 5-20P
100-120 V (1N~) 60 Hz
16 A
CB 160
with O
2
control
9x40-0094
Shock-proof plug
230 V (1N~)
50/60 Hz
16 A
CB 160-UL
with O
2
control
9x40-0095
NEMA 5-20P
100-120 V (1N~) 60 Hz
16 A
CB 220
9x40-0096
Shock-proof plug
230 V (1N~)
50/60 Hz
16 A
CB 220-UL
9x40-0097
NEMA 5-20P
100-120 V (1N~) 60 Hz
16 A
CB 220
with O
2
control
9x40-0098
Shock-proof plug
230 V (1N~)
50/60 Hz
16 A
CB 220-UL
with O
2
control
9x40-0099
NEMA 5-20P
100-120 V (1N~) 60 Hz
16 A
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the unit’s type plate (left unit side, bottom right-hand, see chap. 1.4).
•
When connecting, please observe the regulations specified by the local electricity supply company as
well as the VDE directives (for Germany).
•
Pollution degree (acc. to IEC 61010-1): 2
•
Over-voltage category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 22.4).
To completely separate the unit from the power supply, you must disconnect the power plug.
Install the unit in a way that the power plug is easily accessible and can be easily pulled in
case of danger.