beautypg.com

BINDER CB 60 User Manual

Page 102

background image

CB (E6) 09/2013

Page 102/144

16.1.2 Measuring CO

2

directly via chemical indicator tubes

This is a common “do-it-yourself” test for many users. A chemical color reaction in a glass tube shows the

CO

2

concentration. A standardized volume of air from inside the incubator has to be suctioned through

this glass tube to get a quantitative test result. Therefore, use a special hand pump with a standardized

suction volume.
Procedure (example):
1) Break off both ends of the glass tube or remove the plugs.
2) Pin the end with the higher end of the scale to the adapter of the hand pump that belongs to that test

system.

3) Pin the other end through the silicone access port of the inner chamber door of the CB incubator.
4) Take one sample volume out of the inner chamber volume by pressing the pump fully together and

releasing it afterwards.

5) The standardized volume is suctioned through the glass tube and the chemical indicator changes its

color beginning from the side pinned into the chamber in the direction of the hand pump.

6) The more CO

2

is inside the chamber the more the chemical reaction will cause a color change of the

chemical reactor.

7) You can read the CO

2

concentration by the scale directly printed on the glass tube or a delivered

reference-reading rule.

8) The result must then be corrected to the current ambient pressure. The required formula is printed on

the instruction sheet of such systems.

All the necessary equipment must be supplied by one manufacturer only and in a defined test system.
Note: These test systems are not very accurate. A typical accuracy is around 10 % of the full-scale value.

These test systems are not suitable for calibrating the BINDER sensor system.

Figure 25: Example of chemical indicator

tubes

Figure 26: Example of a hand pump (foreground) and

electrical pump (background)

This manual is related to the following products: