18.3 Hot-air sterilization at 180 °C / 356 °F

The very first sterilization after operation may cause an odor. This is not a quality defect. We

recommend ventilating well the room during sterilization.


18.3.1 Overview

Heating up

Holding

Cooling down

Equilibrating

4 hours

6 hours

unlimited

Sterilization cycle of 10 hours

Figure 27: Setpoint profile during the sterilization cycle



The CB incubator can perform an automatically controlled hot-air sterilization cycle. This procedure will

take approx. 10 hours and consists of the following steps:

Heating up phase: The incubator heats up the inner chamber as fast as possible to the

sterilization setpoint temperature

Holding phase: Constant sterilization setpoint temperature
The sterilization setpoint temperature is pre-set in factory to 187.5 °C / 369.5 °F. The

duration of the heating up and holding phase is in total 4 hours. This ensures that 180 °C /

356 °F is maintained on all internal surfaces for at least 30 minutes

Cooling down phase: Programmed duration of 6 hours until 37 °C / 98.6°F is reached

Equilibration phase: This phase of unlimited duration follows the sterilization with a

constant setpoint temperature of 37 °C / 98.6 °F


To start the hot-air sterilization cycle, go to

Menu > Hot-air sterilization

37 °C

187.5 °C