21.6 Controller

Fault description / indication Possible cause

Required measures

The controller returns to the

initial view from any level..

No button was pressed for

more than 120 sec.

Repeat entries, enter the values

rapidly.

No entries to controller

keypad possible.

Notification

“Key lock”.

Key lock activated.

Enter the key lock password (chap.

8).


21.7 Open door

A door contact switch serves to check the state of the outer door. When the door is open, temperature,

CO

2

, and O

2

(unit with O

2

control) control are turned off.

Fault description / indication Possible cause

Required measures

Outer door open (more

than the selected

alarm delay time)

Alarm message

“Door open”

Outer door open or not properly

closed.

Confirm the alarm (chap. 11.4). Close

the door. The triggered zero-voltage

relay alarm contact switches back.