4 cb technical data, Hap.22.4) – BINDER CB 60 User Manual
Page 131
CB (E6) 09/2013
Page 131/144
Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them
and thus obtain a homogenous distribution of temperature, CO
2
and O
2
(unit with O
2
control).
22.4 CB technical data
Unit size
60
160
220
Exterior dimensions
Width
mm / inch
580 / 22.83
680 / 26.77
740 / 29.13
Height (incl. feet)
mm / inch
720 / 28.35
920 / 36.22
1070 / 42.13
Depth
mm / inch
550 / 21.65
715 / 28.15
715 / 28.15
plus door handle, I-triangle
mm / inch
55 / 2.16
55 / 2.16
55 / 2.16
plus power connection and gas connection
mm / inch
17 / 0.67
17 / 0.67
17 / 0.67
Wall clearance rear (minimum)
mm / inch
100 / 3.94
100 / 3.94
100 / 3.94
Wall clearance side (minimum)
mm / inch
50 / 1.97
50 / 1.97
50 / 1.97
Number of doors
1
1
1
Number of inner glass doors
1
1
1
Interior dimensions
Width
mm / inch
400 / 15.75
500 / 19.69
560 / 22.05
Height
mm / inch
400 / 15.75
600 / 23.62
750 / 29.53
Depth
mm / inch
330 / 12.99
500 / 19.69
500 / 19.69
Interior volume
l / c.ft.
53 / 1.9
150 / 5.4
210 / 7.5
Number of shelves (regular / max.)
2/3
3/6
3/8
Size of shelf (external) Width x Depth
mm x mm
inch x inch
396 x 289
15.59 x 11.38
495 x 444
19.49 x 17.48
556 x 444
21.89 x 17.48
Load per shelf
Kg / lbs.
10 / 22
10 / 22
10 / 22
Permitted total load
Kg / lbs
30 / 66
30 / 66
30 / 66
Weight (empty)
Kg / lbs
60 / 132
107 / 236
121 / 267
Temperature data
Temperature range
from
°C / °F
7 / 12.6
above ambient
7 / 12.6
above ambient
7 / 12.6
above ambient
up to
°C / °F
60 / 140
60 / 140
60 / 140
Temperature fluctuation at 37 °C/ 98.6 °F
± K
0.1
0.1
0.1
Temperature uniformity (variation)
at 37 °C/ 98.6 °F
± K
0.3
0.3
0.4
Recovery time after door was opened for
30 sec at 37 °C/ 98.6 °F
minutes
4
4
5
CO
2
data
CO
2
range
vol.-% CO
2
0 to 20
0 to 20
0 to 20
Setting accuracy
vol.-% CO
2
0.1
0.1
0.1
Recovery time after door was opened for
30 sec at 5 vol.-% CO
2
1)
minutes
5
5
5
CO
2
measurement
IR
IR
IR
Connection hose nozzle DN6 for CO
2
connection to the unit for hose with internal
diameter
mm / inch
6 / 0.24
6 / 0.24
6 / 0.24