Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them

and thus obtain a homogenous distribution of temperature, CO

2

and O

2

(unit with O

2

control).

22.4 CB technical data

Unit size

60

160

220

Exterior dimensions
Width

mm / inch

580 / 22.83

680 / 26.77

740 / 29.13

Height (incl. feet)

mm / inch

720 / 28.35

920 / 36.22

1070 / 42.13

Depth

mm / inch

550 / 21.65

715 / 28.15

715 / 28.15

plus door handle, I-triangle

mm / inch

55 / 2.16

55 / 2.16

55 / 2.16

plus power connection and gas connection

mm / inch

17 / 0.67

17 / 0.67

17 / 0.67

Wall clearance rear (minimum)

mm / inch

100 / 3.94

100 / 3.94

100 / 3.94

Wall clearance side (minimum)

mm / inch

50 / 1.97

50 / 1.97

50 / 1.97

Number of doors

1

1

1

Number of inner glass doors

1

1

1

Interior dimensions
Width

mm / inch

400 / 15.75

500 / 19.69

560 / 22.05

Height

mm / inch

400 / 15.75

600 / 23.62

750 / 29.53

Depth

mm / inch

330 / 12.99

500 / 19.69

500 / 19.69

Interior volume

l / c.ft.

53 / 1.9

150 / 5.4

210 / 7.5

Number of shelves (regular / max.)

2/3

3/6

3/8

Size of shelf (external) Width x Depth

mm x mm

inch x inch

396 x 289

15.59 x 11.38

495 x 444

19.49 x 17.48

556 x 444

21.89 x 17.48

Load per shelf

Kg / lbs.

10 / 22

10 / 22

10 / 22

Permitted total load

Kg / lbs

30 / 66

30 / 66

30 / 66

Weight (empty)

Kg / lbs

60 / 132

107 / 236

121 / 267

Temperature data

Temperature range

from

°C / °F

7 / 12.6

above ambient

7 / 12.6

above ambient

7 / 12.6

above ambient

up to

°C / °F

60 / 140

60 / 140

60 / 140

Temperature fluctuation at 37 °C/ 98.6 °F

± K

0.1

0.1

0.1

Temperature uniformity (variation)

at 37 °C/ 98.6 °F

± K

0.3

0.3

0.4

Recovery time after door was opened for

30 sec at 37 °C/ 98.6 °F

minutes

4

4

5

CO

2

data

CO

2

range

vol.-% CO

2

0 to 20

0 to 20

0 to 20

Setting accuracy

vol.-% CO

2

0.1

0.1

0.1

Recovery time after door was opened for

30 sec at 5 vol.-% CO

2

1)

minutes

5

5

5

CO

2

measurement

IR

IR

IR

Connection hose nozzle DN6 for CO

2

connection to the unit for hose with internal

diameter

mm / inch

6 / 0.24

6 / 0.24

6 / 0.24