4 activating / deactivating all alarm functions – BINDER CB 60 User Manual
Page 80
![background image](https://www.manualsdir.com/files/767051/content/doc080.png)
CB (E6) 09/2013
Page 80/144
11.5.3 Activating, deactivating, and testing the alarm buzzer
Alarm buzzer test
To access the alarm buzzer functional test, go to
Menu > Alarms > Buzzer test
Fixed value
08.03.2013 05:05:06
Submenu “Buzzer test”.
The current setting is displayed.
Press the operating button to turn on or off the
buzzer for test purpose.
The modified setting is displayed.
“Buzzer test: On” = Buzzer turns on
“Buzzer test: Off” = Buzzer turns off
..\ Buzzer test
Buzzer test: Off
Close
Home
When turned on, the alarm buzzer emits an intermittent signal. To turn this off, change the test function to
“Buzzer test: Off”.
Activating / deactivating the alarm buzzer
To activate or deactivate
the alarm buzzer, go to
Menu > Alarms > Buzzer On/Off
Fixed value
08.03.2013 05:05:06
Submenu “Buzzer On/Off”.
The current setting is displayed.
To change the setting, press the operating button.
The modified setting is displayed.
“Buzzer activation: On” = Buzzer will turn on in the
event of an alarm
“Buzzer activation: Off” = Buzzer is deactivated
..\ Buzzer On/Off
Buzzer activation : On
Close
Home
11.5.4 Activating / deactivating all alarm functions
To access the alarm function setting, go to
Menu > Alarms > Alarms On/Off
Fixed value
08.03.2013 05:05:06
Submenu “Alarm functions On/Off”.
The current setting is displayed.
To change the setting, press the operating button.
The modified setting is displayed.
“Alarm functions: On” = The alarm functions of the
controller are active
“Alarm functions: Off” = All alarm functions are
deactivated
..\ Alarm functions On/Off
Alarm functions: On
Close
Home