18.3.2 Performing a hot-air sterilization

Before carrying out the first hot-air sterilization, remove any protective lamination sheet from

the inner metal surfaces.

When starting a hot-air sterilization, all gas controls automatically become inactive.

The safety controller settings are inactive during sterilization. They become functional again

following abortion of the sterilization and/or restart of the unit at the main power switch.

Turn off the unit.

Empty the Permadry™ water pan.

WARNING

Danger of implosion.
Destruction of the incubator.

Empty the Permadry™ water pan before starting the hot-air sterilization.

Water pans and shelves must be inside the incubator.

Close the inner glass door and the outer unit door.

Turn on the unit.

Activate the sterilization procedure in the controller.



To start the hot-air sterilization cycle, go to

Menu > Hot-air sterilization

or

Quick menu > Hot-air sterilization

Fixed value

08.03.2013 05:05:06

Submenu “Hot-air sterilization”.
Select “Start”

and press the operating button.

..\ Hot-air sterilization

Start

Stop

Close