18.3.2 Performing a hot-air sterilization
Before carrying out the first hot-air sterilization, remove any protective lamination sheet from
the inner metal surfaces.
When starting a hot-air sterilization, all gas controls automatically become inactive.
The safety controller settings are inactive during sterilization. They become functional again
following abortion of the sterilization and/or restart of the unit at the main power switch.
•
Turn off the unit.
•
Empty the Permadry™ water pan.
WARNING
Danger of implosion.
Destruction of the incubator.
Empty the Permadry™ water pan before starting the hot-air sterilization.
•
Water pans and shelves must be inside the incubator.
•
Close the inner glass door and the outer unit door.
•
Turn on the unit.
•
Activate the sterilization procedure in the controller.
To start the hot-air sterilization cycle, go to
Menu > Hot-air sterilization
or
Quick menu > Hot-air sterilization
Fixed value
08.03.2013 05:05:06
Submenu “Hot-air sterilization”.
Select “Start”
and press the operating button.
..\ Hot-air sterilization
Start
Stop
Close