3 gas connections, Warning – BINDER CB 60 User Manual
Page 30

CB (E6) 09/2013
Page 30/144
•
We recommend cleaning and refilling the pans 2 to 3 times a week. For evacuation, remove the
Permadry™ water pan.
•
We recommend using distilled, sterile water to achieve optimum growth results. Any corrosive damage
that may arise following the use of water of different quality or by additives is excluded from the liability
agreement.
•
If required, you can add microbiologically inhibiting substances such as copper chips, copper sulfate
or ethylene diamine tetra-vinegar acid (EDTA) in a concentration of 1 to 5 mmol/l.
Empty the Permadry™ water pan before moving the incubator. In case of the contents spilling,
immediately shut down the incubator and dry it carefully and completely.
4.3 Gas connections
General information for safe handling of gas cylinders:
•
Store and use gas cylinders only in well ventilated areas.
•
Open the gas cylinder valve slowly to avoid pressure surges.
•
Secure gas cylinders during storage and use against falling (chaining).
•
Transport gas cylinders with a cylinder cart, do not carry, roll, or throw them.
•
Always close the valve even with apparently empty cylinders; screw on the cap when not in
use. Return gas cylinders with the valve closed.
•
Do not open gas cylinders by force. Mark them when damaged.
•
Protect gas cylinders against fire, e.g. do not store together with flammable liquids.
•
Observe relevant regulations for dealing with gas cylinders.
Secure the gas cylinder against falling and other mechanical damage.
WARNING
Safety valve tearing off.
Sudden release of the stored pressure energy.
Risk of injury.
Secure gas cylinders against falling (chaining).
Transport gas cylinders with a cylinder cart.
The valve of the gas cylinder
always must be closed before screwing on or unscrewing the gas hose.
WARNING
Opening the cylinder valve when the cylinder is not connected.
Sudden release of the stored pressure energy.
Risk of injury.
Close the gas cylinder valve before connecting or removing the gas hose.
After connecting the gas cylinder, check all gas connections for leaks (e.g. with leak spray or
diluted soap solution).