4 confirming an active “set” alarm – BINDER CB 60 User Manual
Page 77
CB (E6) 09/2013
Page 77/144
11.4 Confirming an active “set” alarm
Fixed value
08.03.2013 05:05:06
Initial view with overtemperature safety controller
alarm.
The buzzer sounds (if not deactivated previously).
Press the “Info” button.
Setpoint Actual value
Temperature [°C]
37.0 39.2
CO2 [Vol.-%]
5.0 5.0
Humidity [%]
85.5
User
Info
Menu
Alarm acknowledgement
08.03.2013 05:05:06
Submenu “Alarms” for alarm acknowledgement.
../Home/Alarms
Safety controller overtemperature
Close
Reset
Home
Press the “Reset” button to confirm the alarm. If there is another active “set” alarm, you can also confirm it
with the “Reset” button. After confirming all active “set” alarms, the buzzer is off, and the initial view is
displayed.
As long as an alarm is still active, i.e. as long as the alarm cause is still valid, the alarm icon continues to
be displayed in the initial view.