Page 203

background image

Directions for use

Chapter 11 Specifications

199

Table 43. Review window

OxyCRG

Features

Factory default

HR/PR and SpO

2

trend data

Graphed as lines connecting each successive data point; PR from NIBP
source graphed as single values

Trend data transferred from the patient’s
monitor

Graphed as very small squares

CO

2

and respiration data

Displays as waveforms

Time span

Two minutes

CO

2

or RESP selected

If CO

2

readings were taken in last 24 hours, CO

2

is selected; if no CO

2

readings taken but RESP readings were taken, RESP is selected

Cursor bar position

At right edge of most recent period monitored, or if another review window
was viewed first, the period is centered around the cursor bar position in the
first-viewed review window.

HR/PR and CO

2

labels to the right of the

Graphical Trend display

Reflects label selected at the monitor during the current time position of the
cursor bar

Presets values

Trend scales similar to monitor factory “Trend Graphs” values. For further
information, see the monitor directions for use.

CO

2

range

0 mmHg to 60 mmHg

RESP scale

2x

Blue cursor bar shows

Vital sign numerics

24-hour (or up to 96-hour) time line
indicates patient and equipment state

Time line color matches icon color at all points in time

Printouts

Graph