Directions for use
Chapter 9 Status icons
141
Too Many Hardwired
Monitors
Takes priority
over low-level
alerts.
Location representing the
extra monitor.
[room #] Equipment Audible alert tone
continues until you
disconnect a monitor.
Disconnect a monitor not
in use, or wait until
another port becomes
available to connect the
extra monitor.
Monitor Cannot
Communicate
Take priority
over low-level
alerts
Locations representing
monitors that cannot
communicate with Acuity
System.
[room #] Equipment Close the alert window
and turn off or disconnect
the monitor.
Contact qualified service
personnel.
Monitor Problem
Take priority
over low-level
alerts.
Locations representing
problem monitors.
[room #] Equipment
or more specific
message
At Acuity Central
Station
1. Check yellow Virtual
Monitor for specific
alert message.
2. If there is an audible
tone, click Suspend.
3. If present, press one
of these buttons:
•
DISMISS or
CONFIRM to dismiss
the alert message
•
RESET to reset the
monitor
4. Correct the problem.
If the problem
persists, contact
qualified service
personnel.
At patient monitor
1. Check for specific
alert message.
2. If present, press one
of these buttons:
•
Acknowledge to
acknowledge the
alert
•
RESET to reset the
monitor
•
DISABLE to disable
the alert
3. Correct the problem.
Table 8. Icons and messages that represent patient and monitor status
Patient/monitor
status icon
Icon display
priority
Icon location on map
Messages in
hallway message
panels and at top
of Acuity Central
Station screen
Recommended action